What kind of setting is the item of APPEND of AUTO SAVE setting on LE-series?

OFF: Start measuring after deleting all previous Auto Save data.
ON: Starts measuring not deleting previous Auto Save data.

However, because the number of the files is set by "MAXFILES", if the number of the files is more than the setting, the oldest file will be deleted and the latest files are recorded one by one.

AUTO SAVE performs the ring buffer recording to save the latest file.

If APPEND is ON, it saves previous measurement data to the file as much as possible.